Features
- Integrated Plug in Modular, 12.1” colorful and clear TFT display
 - Standard 6 parameters and optional single plug-in parameter box
 - Touch screen available; able to be operated by keyboard or optional mouse
 - With durable trim knob and backlight silicone buttons
 - Simultaneously display at least 10 parameters, 13 kinds of arrhythmia analysis
 - ECG Waveforms of 7-lead displayed simultaneously.
 - Assist to judge the heart activity directly
 - 15 kinds of drug dose calculations
 - Anti-electrosurgical, anti-defibrillation, suitable for monitoring critical care
 - Trend interface, OxyCRG interface, BigFont interface
 - Built-in detachable rechargeable lithium battery, battery life 4.5 hours
 - 1000 hours trend data and trend graphs, 48 hours full waveform reviewing
 - Power failure data store
 - High capacity of information storage
 - NIBP dual overpressure protection
 - Wired/wireless connection, with system supporting 3 display screens
 - Supporting SD card, 4 USB interfaces
 - Omni-directional visible design; 3 levels of audible, visual and audio alarm; dual alarm lights for physiological and technical alarm
 
Technical specification
ECG
- Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
 - Lead selection: I, II, III, avR, avL, avF, V,
 - Waveform: 2 ch
 - Lead mode: 3 Leads (R, L, F or RA, LA, LL)
 - Lead selection: I, II, III,
 - Waveform: 1 ch
 - Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
 
Heart Rate
- Range
- Adult: 15 ~ 300 bpm
 - Neo/Ped: 15 ~ 350 bpm
 
 - Accuracy: ± 1% or ± 1bpm, which great
 - Resolution: 1 bpm
 - Sensitivity: > 200 (uV P-P)
 - Differential Input Impedance: > 5 MΩ
 - CMRR
- Monitor: > 105 dB
 - Operation: > 105 dB
 - Diagnosis: > 85 dB
 
 - Electrode offset potential: ±300mV
 - Leakage Current: < 10 uA
 - Baseline Recovery: < 3 S after Defi.
 - ECG Signal Range: ±8 m V (Vp-p)
 - Bandwidth
- Surgery: 1 ~ 15 Hz
 - Monitor: 0.5 ~ 35 Hz
 - Diagnostic: 0.05 ~ 100 Hz
 
 - Calibration Signal: 1 (mV p-p), Accuracy: ±5%
 - ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
 - ARR Detecting
 - Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
 - Alarm: Available
 - Review: Available
 
Respiration
- Method: Impedance between R-F (RA-LL)
 - Differential Input Impedance: >2.5 MΩ
 - Measuring Impedance Range: 0.3~5.0Ω
 - Base line Impedance Range: 0 – 2.5 KΩ
 - Bandwidth: 0.3 ~ 2.5 Hz
 - Resp. Rate
 - Measuring and Alarm Range
- Adult: 0 ~ 120 rpm
 - Neo/Ped: 0 ~ 150 rpm
 
 - Resolution: 1 rpm
 - Accuracy: ±2 rpm
 - Apnea Alarm: 10 ~ 40 S
 
NIBP
- Method: Oscillometric
 - Mode: Manual, Auto, STAT
 - Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
 - 120, 180, 240,480 (Min)
 - Measuring Period in STAT Mode: 5 Min
 - Pulse Rate Range: 40 ~ 240 bpm
 - Alarm Type: SYS, DIA, MEAN
 - Measuring and alarm range
 -   Adult Mode
- SYS: 40 ~ 270 mmHg
 - DIA: 10 ~ 215 mmHg
 - MEAN: 20 ~ 235 mmHg
 
 -   Pediatric Mode
- SYS: 40 ~ 200 mmHg
 - DIA: 10 ~ 150 mmHg
 - MEAN: 20 ~ 165 mmHg
 
 - Neonatal Mode
- SYS: 40 ~ 135 mmHg
 - DIA: 10 ~ 100 mmHg
 - MEAN: 20 ~ 110 mmHg
 
 - Resolution Pressure: 1mmHg
 - Accuracy Pressure Maximum Mean error: ±5mmHg
 - Maximum Standard deviation: ±8mmHg
 - Overpressure Protection
- Adult Mode: 297±3 mmHg
 - Pediatric Mode: 240±3 mmHg
 - Neonatal Mode: 147±3 mmHg
 
 
SpO2
- Measuring Range: 0 ~ 100 %
 - Alarm Range: 0 ~ 100 %
 - Resolution: 1 %
 - Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified
 - Actualization interval: about 1 Sec.
 - Alarm Delay: 10 Sec.
 
Pulse Rate
- Measuring and Alarm Range: 20~300bpm
 - Resolution: 1bpm
 - Accuracy: ±2bpm
 
Temperature
- Channel: 2
 - Measuring and Alarm Range: 0 ~ 50 °C
 - Resolution: 0.1°C
 - Accuracy: ±0.1°C
 - Actualization interval: about 1 Sec.
 - Average Time Constant:<10 Sec.
 
Configuration
- Standard: ECG, RESP, NIBP, SPO2, PR, TEMP
 - Optional Function: Recorder, 2-IBP, VGA, ETCO2, AG (Anesthesia Gas), CO (Cardiac Output), Touch Screen, WIFI, Central Monitoring System.
 