Specifications
| Model | AC9600 |
| Measuring item | K,Na,Cl,TCO2 |
| Calculating item | AG |
| Measuring time | Measuring time: 55s Time for sampling, measuring, washing and printing: 90s |
| Sample size | 160ul |
| Data storage | 10000 |
| Communication interface | 232 interface |
| Automatic sampling | 25-bit sampling system for option (20 samples, 2 QC, 1 emergency , 1 washing) |
| Air bubble test | Yes |
| Alarm for waste liquid overflow | Yes |
| Alarm for standard liquid level | Yes |
| Standard liquid level indicator lamp | Yes |
| Sample liquid level indicator lamp | Yes |
| Power supply | AC220V±22V 50Hz 50W |
| Size | 408mmX390mmX230mm |
| Weight | 7Kg(without sampling system)1Kg (sampling system) |
| Measuring method | ISE. TCO2 adopts pressure-measuring method. |
| Working condition | Temperature: 10℃~32℃ Relative humidity ≤85% |
| Â Measuring Parameter and Specification Table |
| Measuring item | Measuring scope | Resolving power | Precision |
| (K+) | 0.50—15.00 | 0.01 | CV≤1.0% |
| (Na+) | 30.0—200.0 | 0.1 | CV≤1.0% |
| (Cl-) | 30.0—200.0 | 0.1 | CV≤1.0% |
| (Ca2+) | ——- | ——– | ——- |
| (H+) | ——- | ——- | ——- |
| (TCO2) | 4.0—70.0 | 0.1 | CV≤3.0% |