Technical Specifications
PARAMETER | SPECIFICATION |
System function | |
General | Automatic, discrete, random access
STAT sample priority |
Throughput | Constant 240T/H, up to 300T/H |
Principle | Colorimetry, turbidimetry |
Methodology | End-point, fixed-time and kinetic |
Calibration | Linear, Logit-Log 4P, Logit-Log 5P,
Spline, Exponential, Polynomial |
Reagent/Sample handling | |
Reagent/Sample tray | 60 reagent positions
60 sample positions in cooling compartment (2 to 12°C) |
Barcode scanner | Sample barcode scanner (optional) |
Reagent volume | 20 to 450 ul, step by 0.1 ul |
Sample volume | 2 to 60 ul, step by 0.1 ul |
Reagent/Sample probe | Liquid level detection and tracking
Collision protection Inventory checking |
Probe cleaning | Automatic interior and exterior wash |
Automatic dilution | Pre-dilution and post-dilution
Dilution ratio up to 1:200 |
Reaction system | |
Cuvette | 81 resusable cuvettes |
Reaction volume | 150 to 500 ul |
Reaction temperature | 37.0 ± 0.1°C |
Mixer | Independent mixing bar |
Cuvette washing | Auto washing station |
Optical system | |
Light source | Halogen tungsten lamp |
Photometer | Maintenance-free sealed photometer |
Wavelength | 8 wavelengths: 340nm, 405nm, 450nm,
510nm, 546nm, 578nm, 630nm, 670nm |
Maximum Abs | up to 4.5 Abs, relative bias ≤ ± 2% |
Working Condition | |
Operation system | Windows 7/8/10 |
Connection to PC | Network port |
PC | Compatible with wide-screen |
Power supply | AC 100~240V, 50/60Hz, ≤ 450VA |
Temperature | 10-30°C |
Humidity | 30-85% RH |
Water consumption | ≤ 8.4 L/H |
Dimension | 880mm (W) x650mm (D) x570mm (H) |
Reviews