II. Specification
Item | Parameter | Remark |
X-ray Generator | Generator Type: High Frequency Inverter 80kHz
Input Power: Single phase 220VAC, 50/60Hz Radiographic Ratings: Large Focal Point 20-35kV/10-510mAs Small Focal Point 20-35kV/10-100mAs Power Rating: 6.2kVA |
Self-developed and world advanced all-solid-state high frequency high voltage x-ray generator |
X-ray Tube | Focal Spot Size: Dual Focus 0.1/ 0.3mm
Target Material: Molybdenum (Mo) Port Material: Beryllium (Be) High-speed anode drive: 2800 /10000rpm Target angle:10°/16° Anode Heat Storage: 210kJ (300kHU) Anode Cooling: Air cooling Filtration: Mo(0.03mm), Al(0.5mm) |
Model:IAE C339V
China tube for optional |
Radiographic Stand | C-ARM: Vertical Movement: 590mm
Center of electricrotating C-arm Automaticreturn function by one key Rotations Degree: +90°~-90° Automatically released after the exposure pressure settingsdisplay Compression flexible, stepless speed. Max. pressure: 200N Max. travel: 150mm SID: 650mm |
Electric Isocentricrotating |
Flat Panel Detector | Detector material: Amorphous silicon
Effective coverage of detector: 18x24cm Pixel matrix: 3072×1944 Limit of spatial resolution: 6.0Lp/mm DQE value: 70% dynamic range: 14bit digital output pixel size: 75μm High voltage Synchronizer trigger: BNC Output: Camera Link or Ethernet Working condition: 10℃-40℃ storage environment: -10℃-50℃ |
China Flat Panel Detector
24x30cm for optional |
Bucky housing and movement device | Size: 374*304*65mm
Stepless speed regulating range: 0~6cm/s Movement range: 0.5~2cm Grid Size: 24x30cm Grid ratio: 5:1 Grid density: 30lp/cm Focal distance: 650mm |
|
Image acquisition workstation | CPU≥Intel Core Duo 2.60GHz
Hardware≥250G high speed Hardware Memory≥2G Display card≥512MB high brightness high-contrast LCD,1280*1024 Pixel resolution Network interface Work-list DICOM3.0 transmission 100/1000 Gigabit Ethernet Software Imaging software packageDMOC V1.0 |
Configuration including Diagnose digital workstation
5M medical monitor for optional |
Others | Line Voltage
220Vac±10%@25A,Single phase |
110V for optional |
III. Configuration
No. | Item | Quantity |
1 | X-ray Tube | 1 |
2 | X-ray Generator | 1 |
3 | Gantry assembly | 1 |
4 | C-ARM | 1 |
5 | Bucky movement device | 1 |
6 | Flat panel detector | 1 |
7 | Image acquisition workstation | 1 |
8 | Review work station | 1 |
9 | Paddle switch | 2 |
10 | Exposal switch and connected line | 1 |
11 | Power wire | 1 |
12 | Grounded wire | 1 |
13 | Fuse | 2 |
14 | Operation manual | 1 |
15 | Maintenance Reference Manual | 1 |
- IV. Features
- Adopt specialized mammography flat panel detector digital imaging technology.
- Full size digital mammography x-ray imaging.
3. Unique adopt all-solid-state high frequency high voltage generator. This technology has got the PATENT IN THE USA.
Reviews