Specifications
- Sterilizing course automatic sterilization controlled by computer, easy to operate.
 - With drying function, suitable dressing drying.
 - With over-temperature, over-pressure auto-protect device.
 - The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.
 - It can’t be start on if the door doesn’t close well.
 - The safety valve will be open automatically when the inner pressure over 0.24MPa, and the steam be exhausting to the water tank.
 - The power will be cut -off automatically, cut-off the water and alarm when the machine lacking water.
 
Technical Data
| Model technical data | WS-400YDA | WS-500YDA | 
| Volume of sterilizing chamber | 400L, Φ700×1100mm | 500L, Φ700×1300mm | 
| working pressure | 0.22MPa | |
| working temperature | 134℃ | |
| Adjustment of temperature | 105℃ – 134℃ | |
| Time of sterilization | 0-60 min | |
| Time of drying | 0-60 min | |
| Heat average | ≤±2℃ | |
| Power Supply with mains steam | 18KW/380V 50Hz | 18KW/380V 50Hz | 
| Dimension | 1430×880×1830 | 1800×900×1820 | 
| Transport size | 1600×1050×2100 | 1850×1050×2100 | 
| Weight | 530/420 KG | 580/470 KG | 