Technical Specification
ECG
- Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
 - Lead selection: I, II, III, avR, avL, avF, V,
 - Waveform: 2 ch
 - Lead mode: 3 Leads (R, L, F or RA, LA, LL)
 - Lead selection: I, II, III,
 - Waveform: 1 ch
 - Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
 
Heart Rate and Alarm
- Range
 - Adult: 15 ~ 300 bpm
 - Neo/Ped: 15 ~ 350 bpm
 - Accuracy: ± 1% or ± 1bpm, which great
 - Resolution: 1 bpm
 - Sensitivity: > 200 (uV P-P)
 - Differential Input Impedance: > 5 MΩ
 - CMRR
- Monitor: > 105 dB
 - Operation: > 105 dB
 - Diagnosis: > 85 dB
 
 - Electrode offset potential: ±300mV
 - Leakage Current: < 10 uA
 - Baseline Recovery: < 3 S after Defi.
 - ECG Signal Range: ±8 m V (Vp-p)
 - Bandwidth
- Surgery: 1 ~ 15 Hz
 - Monitor: 0.5 ~ 35 Hz
 - Diagnostic: 0.05 ~ 100 Hz
 
 - Calibration Signal: 1 (mV p-p), Accuracy: ±5%
 - ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
 - ARR Detecting
 - Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
 - Alarm: Available
 - Review: Available
 
Respiration
- Method: Impedance between R-F (RA-LL)
 - Differential Input Impedance: >2.5 MΩ
 - Measuring Impedance Range: 0.3~5.0Ω
 - Base line Impedance Range: 0 – 2.5 KΩ
 - Bandwidth: 0.3 ~ 2.5 Hz
 - Resp. Rate
 - Measuring and Alarm Range
 - -Adult: 0 ~ 120 rpm
 - -Neo/Ped: 0 ~ 150 rpm
 - Resolution: 1 rpm
 - Accuracy: ±2 rpm
 - Apnea Alarm: 10 ~ 40 S
 
NIBP
- Method: Oscillometric
 - Mode: Manual, Auto, STAT
 - Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
 - 120, 180, 240,480 (Min)
 - Measuring Period in STAT Mode: 5 Min
 - Pulse Rate Range: 40 ~ 240 bpm
 - Alarm Type: SYS, DIA, MEAN
 
Measuring and alarm range
Adult Mode
- SYS: 40 ~ 270 mmHg
 - DIA: 10 ~ 215 mmHg
 - MEAN: 20 ~ 235 mmHg
 
Pediatric Mode
- SYS: 40 ~ 200 mmHg
 - DIA: 10 ~ 150 mmHg
 - MEAN: 20 ~ 165 mmHg
 
Neonatal Mode
- SYS: 40 ~ 135 mmHg
 - DIA: 10 ~ 100 mmHg
 - MEAN: 20 ~ 110 mmHg
 - Resolution Pressure: 1mmHg
 - Accuracy Pressure Maximum Mean error: ±5mmHg
 - Maximum Standard deviation: ±8mmHg
 - Overpressure Protection
- Adult Mode: 297±3 mmHg
 - Pediatric Mode: 240±3 mmHg
 - Neonatal Mode: 147±3 mmHg
 
 
SpO2
- Measuring Range: 0 ~ 100 %
 - Alarm Range: 0 ~ 100 %
 - Resolution: 1 %
 - Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified
 - Actualization interval: about 1 Sec.
 - Alarm Delay: 10 Sec.
 - Pulse Rate
 - Measuring and Alarm Range: 20~300bpm
 - Resolution: 1bpm
 - Accuracy: ±2bpm
 
Temperature
- Channel: 2
 - Measuring and Alarm Range: 0 ~ 50 °C
 - Resolution: 0.1°C
 - Accuracy: ±0.1°C
 - Actualization interval: about 1 Sec.
 - Average Time Constant:<10 Sec.
 
IBP(optional)
- Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
 - Measuring and alarm range
- ART: 0 ~ 300 mmHg
 - PA: -6 ~ 120 mmHg
 - CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
 - P1/P2: -10 ~ 300 mmHg
 
 - Press Sensor
 - Sensitivity: 5 uV/V/mmHg
 - Impedance: 300-3000Ω
 - Resolution: 1 mmHg
 - Accuracy: ±2% or ±1mmHg, which great
 - Actualization interval: about 1 Sec
 
Standard Packing List
- 5-lead ECG cable (including RESP)
 - Adult finger SpO2 sensor
 - Adult NIBP cuff
 - NIBP extension tube
 - TEMP sensor
 - AC power adaptor
 - ECG electrodes
 - Operation instruction
 
Optional Function
Internal Recorder, IBP, EtCO2, Central Monitoring System, Touch Screen, VGA, Neonate SPO2