- Using the world’s top high-end flat panel detector, high resolution,
small distortion, uniform brightness.
- Use exposure mode, only when needed to activate the X source. During the rotation of the 18s, the rays last only 4-8s. The effect of reducing the check dose, reached the international advanced level.
- 3000A proprietary three-dimensional reconstruction algorithm, the original two-dimensional sequence projection into a three-dimensional volume image, you can at any angle, any location to provide high-definition tomography.
- through the three-dimensional volume image extraction of oral panorama, can provide high-definition panoramic images.
5, the image can be used to generate reports, print or save as a report file.
6, can provide standing, sitting and wheelchair and other kinds of film.
| III. Standard Configuration | |
| 1、Main Frame | 1 set |
| 2、High Frequency High Voltage X-ray Generator | 1 set |
| 3、19 inch color ordinary LCD monitor | 1 set |
| 4、12cmX15cm flat panel detector | 1 set |
| 5、Digital Acquisition Processing Workstation | 1 set |
| 6、Electrically adjustable collimator | 1 set |
| 7、Graphical LCD touch screen | 1 unit |
| 8、Wall-mounted control box | 1 unit |
| 9、Electrical control cabinet | 1 unit |
| 10、Printer (optional) | 1 set |
- Technical Specifications
| PARAMETER | SPECIFICATION | |
| Technical indicators | Anode voltage: 60-92 kV | |
| anode current: 1-15 mA. | ||
| Maximum output power: 1.38kw | ||
| parameter adjustment range: 60 ~ 92KV 1 ~ 15mA | ||
| Information display: KV, MA, lifting position information, human characteristics, laser indication and positioning | ||
| X-ray tube | X-ray Tube Model: D-054SB (Toshiba) | |
| Focus: 0.5 | ||
| Heat capacity: 35kJ (50kHU) | ||
| photography mode: cone beam CT | ||
| Image system Thales flat panel detector 650HD-E | image receiving size: 12cmX15cm flat panel detector | |
| pixel size: 150 | ||
| A / D conversion depth: 14bit | ||
| Acquisition mode: pulse mode | ||
| Detector Type: CMOS | ||
| Pixel matrix: 960 * 786 | ||
| Limit resolution up to: two-dimensional ≥ 3.1 lp / mm | ||
| Structural Performance | Column lift stroke: 1000mm | |
| Lifting noise: <70dB (with buzzer sound) | ||
| Collision protection: Yes | ||
| Weight: 260kg | ||
| Dimensions: 788mm (width) X 1090mm (length) X 2200mm (height) | ||
| Power requirements: 220V / 50HZ / 2kVA | ||
| Space requirements: the required area of 5 square meters, unilateral 2 meters | ||
| The camera room should have a protective thickness of 2 mm lead | ||
| Electrical control device size: width 340mm, depth 550mm, height 650mm |